Sign In | Join Free | My ledscreensign.com
ledscreensign.com
Products
Search by Category
Home > Instrument Parts & Accessories >

30-130 Fully Compensated Temperature And Pressure Output Medical Pressure Sensors

Categories Medical Pressure Sensors
Brand Name: Consensic
Model Number: CPS121
MOQ: Please contact us for details
Price: Offline contact or face-to-face discussion can be negotiated
Payment Terms: L/C/Ade/Pellet/Texi Union, MoneyGram, Alipay, and business-to-business payments
Supply Ability: 999
Delivery Time: 3-7 work days
Place of Origin: WUXI
Certification: PTC21051702401C-EN01
Pressure accuracy: <±0.3kpa, 700kpa, 1MPa, 1.5MPa
type of output: I2C output/analog output
size: 3x5x1.2mm
Minimum number of packages: 10pcs
encapsulation: LGA
traits: Low power consumption
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

30-130 Fully Compensated Temperature And Pressure Output Medical Pressure Sensors


CPS121 Accuracy<± 0.3kpa, 30-130 fully compensated temperature and pressure output

Product Description


The CPS121 system-in-a-package (SIP) solution comprises of aresistive bridge type pressure sensor and a 24-bit ADC for high resolution and accurate pressure measurements. The fully calibrated pressure and temperature compensated digital output makes the CPS121 solution simple to use.The CPS121includes internal calibration logic that provides accurate pressure and temperature measurements to the application via the I²C interface.There is no need to separately download internal calibration coefficients and have the host microcontroller perform complicated compensation calculations.


Features


  • Factory Calibrated Pressure and Temperature Sensor
  • Supply Voltage: 2V to 5.5V(3V @typical)
  • Average Current Consumption: <5uA (One Measurement)
  • Sleep State Current Consumption: <200nA (25°C)
  • Operating Temperature Range: -40°C to +85°C
  • Pressure Absolute Accuracy: ±0.1kPa (±1.0mbar) @ 0°Cto 50°C, 95kPa to 105kPa
  • Pressure Relative Accuracy: ±0.01kPa(<1m)
  • Temperature Accuracy: ±1.0°C

China

Wuxi Consensic Electronics Co., Ltd.

No.86,Dicui Road,B5 Building,

Fourth Floor,Room 401,

Wuxi, Jiangsu Province, 214072

Ph: +86 510.85122279

Fax: +86 510.85122259


Operating Characteristics


Absolute Ratings

PARAMETERSYMBOLCONDITIONSMINTYPMAXUNITS
Over Pressure2X FS

kPa

(bar)

Supply Voltage

(with respect to GND)

VDD-0.36.5V

Voltages at Analog and

Digital I/O Pins

VA_IO

VD_IO

-0.3VDD +0.3V
Storage TemperatureTSTOR-60150°C

Operating Conditions

PARAMETERSYMBOLCONDITIONSMINTYPMAXUNITS
PRESSURE SENSOR
Range

30

(300)

120

(1200)

kPa

(mbar)

Resolution10.17Pa
Noise in Pressure

Full Bandwidth, Normal Mode

Altitude Based on Relative Pressure

1

10

Pa

cm

Accuracy

30 to 120kPa

(-20°C to 0°C)

-0.2

(-2.0)

±0.17

(±1.7)

+0.2

(+2.0)

kPa

(mbar)

30 to 120kPa

(0°C to 65°C )

-0.15

(-1.5)

±0.10

(±1.0)

+0.12

(+1.2)

kPa

(mbar)

Solder Drifts-0.1+0.2kPa
TEMPERATURE SENSOR
Range-4085°C
Resolution0.003°C
Accuracy-40°C to 85°C-1±0.75+1°C
OPERATION
Supply Voltage to GND2VSUPPLY23.05.5V
Operating Temperature Range-4085°C
I2C Pull-Up ResistorsRPU12.2kΩ

1Guaranteed by design of 24bits ADC, and calculated according to the range in application.

2 Factory calibrated for Pressure and Temperature at 3.0V±10%. Output accuracy will be affected if used outside this range. Other ranges available upon request.


Electrical Parameters

PARAMETERSYMBOLCONDITIONSMINTYPMAXUNITS
SUPPLY CURRENT

Supply Current,

average1

during conversion2

standby(no conversion)


Iavg

Isc

Iss


VDD=3V



3.95

1.5



0.2


µA

mA

µA

ANALOG TO DIGITAL CONVERTER
ResolutionrADC24Bit
I2C Clock FrequencyFC,I2C400kHz

1Under the assumption of one conversion every second. Conversion means either a pressure or a temperature measurement

2During conversion, the sensor will be switched on to VDD, and after conversion ended, the sensor will automatically be switched off from VDD.

Buy 30-130 Fully Compensated Temperature And Pressure Output Medical Pressure Sensors at wholesale prices
Send your message to this supplier
 
*From:
*To: Wuxi Consensic Electronic Technology Co., Ltd.
*Subject:
*Message:
Characters Remaining: (0/3000)
 
Find Similar Products By Category:
Inquiry Cart 0