The KTG112-W Trace Oxygen Analyzer Is Applied To The On-line Analysis Of Nitrogen Production From Air Separation.

Categories Gas Detector Sensor
Brand Name: KACISE
Model Number: KTG112-W
Certification: CE
Place of Origin: CHINA
MOQ: 1PCS
Payment Terms: L/C, D/A, D/P, T/T, Western Union, MoneyGram
Supply Ability: 1000 Piece/Pieces per Week negotiable
Delivery Time: 5-8 work days
Packaging Details: each unit has individual box and all boxes are packed in standard packages or customers requests available
Measurement Range: (0.0 - 10/100/1000) × 10⁻⁶;
Maximum Allowable Error: ±2% FS (0.0 - 10) × 10⁻⁶; ±1% FS (10 - 1000) × 10⁻⁶;
Weight: Approximately 1.5Kg.
Repeatability: ±1% FS;
Sample Gas Flow Rate: (400 ± 20) mL/min;
Response Time: τ90 ≤ 60s;
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

The KTG112-W Trace Oxygen Analyzer Is Applied To The On-line Analysis Of Nitrogen Production From Air Separation.

The KTG112-W trace oxygen analyzer is applied to the on-line analysis of nitrogen production from air separation


Introduction

The KTG112-W trace oxygen analyzer is a new type of intelligent industrial portable analyzer developed by combining the electrochemical detection principle, imported high-performance electrochemical sensors, and new microcomputer technology.

Application Areas

Portable and rapid analysis of trace oxygen in nitrogen production from air separation, chemical processes, protective gases in high-temperature sintering furnaces for magnetic materials, protective gases in the electronics industry, as well as in the glass and building materials industries and various mixed gases.


Instrument Features

  • Selects imported fuel cell sensors, featuring long service life, high accuracy, and fast response, and is not interfered with by other gas components.
  • Has automatic temperature compensation to eliminate the influence of ambient temperature.
  • Optional components such as precision filters and air pumps can be added.
  • Uses a 128×64 dot-matrix OLED liquid crystal display, which can simultaneously show parameters such as oxygen content, date, time, and the internal temperature of the instrument.
  • Has a timed automatic storage function, allowing users to view stored data at any time.
  • Alarm points can be set arbitrarily within the full measurement range.
  • Can be powered by both AC and DC, offering flexible use. The battery can work continuously for about 10 hours after a full charge, and it will automatically alarm when the battery is under - voltage.

Technical Parameters

Measurement Range: (0.0 - 10/100/1000) × 10⁻⁶;

Maximum Allowable Error: ±2% FS (0.0 - 10) × 10⁻⁶; ±1% FS (10 - 1000) × 10⁻⁶;

Stability: Zero drift: ±1% FS/7d; Span drift: ±1% FS/7d;

Repeatability: ±1% FS;

Sample Gas Flow Rate: (400 ± 20) mL/min;

Response Time: τ90 ≤ 60s;

Resolution: 0.1 × 10⁻⁶;

Sample Gas Pressure: 0.05 MPa ≤ Inlet pressure ≤ 0.25 MPa;

Operating Environment: Temperature: -15°C - +45°C; Humidity: ≤90% RH;

Operating Power Supply: Rechargeable battery built - in the instrument;

Charging Power Supply: (220 ± 22) VAC, (50 ± 5) Hz. The charger has a built - in charging protection function;

Overall Dimensions: 255mm (Width) × 103mm (Height) × 265mm (Depth)